CVE-2020-12041 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to…

Description

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Affected products and versions

Product Affected versions Fixed in
Baxter Sigma Spectrum Infusion Pumps Versions not published
The 6 CVEs of Baxter Sigma Spectrum Infusion Pumps →