Not rated
CVE-2020-12041 The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to…
Description
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
Affected products and versions
| Product | Affected versions | Fixed in |
|---|---|---|
| Baxter Sigma Spectrum Infusion Pumps | Versions not published |
Other CVEs of Baxter Sigma Spectrum Infusion Pumps
The 6 CVEs of Baxter Sigma Spectrum Infusion Pumps →
CVE-2020-12040
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model…
Not rated
CVE-2020-12039
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion…
Not rated
CVE-2020-12045
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x…
Not rated
CVE-2020-12043
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP…
Not rated